Product Identification

  • Manufacturer: PCB Piezotronics, Inc.

  • Model: 356A32/NC.

  • Product Type: Triaxial ICP® Accelerometer.

  • Sensing Technology: Piezoelectric.

  • Signal Conditioning Technology: ICP® / IEPE.

  • Measurement Axes: X, Y and Z.

  • Ordering Configuration: No mating cable supplied.

Performance Specifications

  • Nominal Sensitivity: 100 mV/g.

  • SI Sensitivity: 10.2 mV/(m/s²).

  • Sensitivity Tolerance: ±10%.

  • Measurement Range: ±50 g peak.

  • SI Measurement Range: ±491 m/s² peak.

  • Frequency Range (±5%): 1–4,000 Hz.

  • Frequency Range (±10%): 0.7–5,000 Hz.

  • Resonant Frequency: ≥25 kHz.

  • Typical Broadband Resolution (1–10,000 Hz): 0.0003 g RMS.

  • SI Broadband Resolution: 0.003 m/s² RMS.

  • Nonlinearity: ≤1%.

  • Transverse Sensitivity: ≤5%.

Electrical Specifications

  • Excitation Voltage: 24–30 VDC.

  • Constant Current Excitation: 2–20 mA.

  • Output Impedance: ≤200 Ω.

  • Output Bias Voltage: 7–16 VDC.

  • Discharge Time Constant: 0.4–1.5 seconds.

  • Settling Time Within 10% of Bias: Less than 5 seconds.

  • Measurement Output: Dynamic analog voltage.

Typical Spectral Noise

  • At 1 Hz: 95 µg/√Hz.

  • At 10 Hz: 20 µg/√Hz.

  • At 100 Hz: 6 µg/√Hz.

  • At 1 kHz: 3 µg/√Hz.

  • At 10 kHz: 2 µg/√Hz.

Physical Specifications

  • Sensing Element Material: Ceramic.

  • Sensing Geometry: Shear.

  • Housing Material: Titanium.

  • Sealing: Hermetic.

  • Dimensions (H × L × W): 11.4 × 11.4 × 11.4 mm.

  • Typical Weight: 5.4 g.

  • Electrical Connector: Mini 8-36, four-pin.

  • Electrical Connection Position: Side.

  • Mounting Thread: 5-40 Female.

  • Mounting Torque: 0.45–0.55 Nm.

Environmental Specifications

  • Operating Temperature Range: -54°C to +121°C.

  • Shock Overload Limit: ±5,000 g peak.

  • SI Shock Overload Limit: ±49,050 m/s² peak.

  • Typical Base Strain Sensitivity: 0.001 g/µε.

  • SI Base Strain Sensitivity: 0.01 (m/s²)/µε.

Typical Applications

  • Industrial machinery vibration testing.

  • Motor and pump vibration analysis.

  • Mechanical structural testing.

  • Experimental modal analysis.

  • Automotive NVH testing.

  • Product vibration qualification.

  • Laboratory dynamic measurements.

  • Multichannel vibration data acquisition.

Important Technical Notes

The 356A32/NC requires compatible constant-current ICP®/IEPE excitation and does not provide an integrated 4–20mA industrial transmitter output.

The shock overload limit of ±5,000g is not the normal acceleration measurement range.

The /NC configuration does not include a mating cable. Cable selection, mounting accessories and calibration documentation must be confirmed before ordering.

All specifications apply at room temperature unless otherwise specified.